- Know the device: pacemaker or ICD? Leads or leadless? Pacemaker-dependent? Card, CXR, ECG, or the device clinic.
- Magnet on a pacemaker → asynchronous pacing (VOO/DOO). Magnet on an ICD → shocks/ATP off, pacing unchanged. Most leadless pacemakers ignore magnets.
- One ICD shock, feels fine → interrogation within days. ≥2 shocks, storm, syncope, chest pain → ED/telemetry, K/Mg, ischemia check, interrogate now.
- Inappropriate shocks (AF, noise, lead fracture) → magnet to stop shocks while you get EP — on telemetry with external defibrillator ready.
- Before cardioversion/defibrillation: pads ≥8 cm from the generator (anterior–posterior). Interrogate after.
- End of life: offer to turn off ICD shocks — a patient's right; it is not euthanasia.
Pathophysiology in 60 seconds
⌄
- Pacemaker: senses the heart's own beats; if none comes in time, it paces. Leads in the RA and/or RV (or a capsule in the RV/RA for leadless).
- ICD: a pacemaker that also detects VT/VF → ATP (fast pacing to break VT painlessly) → shock. S-ICD (subcutaneous) shocks only, no routine pacing.
- CRT: a third lead (LV via coronary sinus) or conduction-system pacing → both ventricles squeeze together → better EF in LBBB.
- RV-only pacing activates the heart like LBBB → long-term can cause pacing-induced cardiomyopathy (high pacing burden).
Frontal view, as you face the patient (patient's right is on your left). Order of the impulse: SA node → atria (Bachmann's bundle to the LA) → AV node (slow, makes the PR) → His bundle (crosses the fibrous skeleton) → right bundle (thin cord down the septum, then the moderator band) and left bundle (short trunk that fans into the anterior and posterior fascicles) → Purkinje fibers. Diagram adapted from the jPoint EP Lab model.
Diagnosis — reading the device
⌄
DDD = paces and senses both chambers, inhibits or triggers. VVI = paces and senses the ventricle, inhibits if it sees a native beat. Magnet response differs by vendor and programming — confirm with the device clinic.
| Problem | ECG | Common causes |
|---|---|---|
| Failure to capture | Spike, no QRS/P after it | Lead dislodgement (first weeks), high threshold (hyperkalemia, acidosis, ischemia, flecainide), lead fracture |
| Failure to output | Missing spikes when they should come | Oversensing (noise, EMI, T waves), battery depletion, lead fracture |
| Undersensing | Spikes on top of native beats, at random | Low signal (new MI, lead issue) — risk of spike on T → VF (rare) |
| Pacemaker-mediated tachycardia | Paced at upper rate limit, regular | Retrograde P triggers ventricular pacing loop → magnet stops it |
A pacemaker is not a defibrillator. A patient with a pacemaker and VT/VF still needs ACLS. Do CPR and shock normally — just keep pads away from the generator.
Workup
⌄
- Interrogation (device rep / EP clinic / remote monitoring): battery, lead impedance, thresholds, events, % pacing, stored VT/VF and shocks.
- ECG with and without magnet (pacemaker) — shows capture at the magnet rate and battery status (magnet rate drops when ERI).
- CXR (PA + lateral): lead position, fracture, pneumothorax (new implant).
- BMP (K), Mg, troponin (shocks), TSH, drug levels (antiarrhythmics, digoxin).
- Device infection signs (pocket redness, erosion, bacteremia, especially S. aureus) → blood cultures ×2–3, TEE → see Endocarditis.
Management
⌄
| Scenario | Action |
|---|---|
| Single shock, now well | Interrogation in 24–72 h, check K/Mg, meds |
| ≥2 shocks in 24 h / storm (≥3 VT/VF in 24 h) | Admit, telemetry, interrogate, fix K/Mg, ischemia work-up, amiodarone + β-blocker (propranolol), sedation — see VT chapter |
| Inappropriate shocks (AF with RVR, sinus tach, noise) | Magnet to suspend, rate-control AF, EP to reprogram/fix the lead |
| Shock + syncope or chest pain | ED, ACS rule-out |
- Above the umbilicus with monopolar cautery → EMI risk is higher. Pacemaker-dependent patient: magnet (asynchronous) or reprogram. ICD: magnet or reprogram to suspend shocks.
- Below the umbilicus, bipolar cautery → often no change needed.
- Always: external pads on, continuous monitoring, restore and interrogate after if reprogrammed or cautery was used near the device.
- Prone position or device in the surgical field → reprogram, a magnet may not stay put.
- MRI: most current systems are MR-conditional → device team programs MRI mode. Legacy (non-conditional) systems can still be scanned under a protocol at experienced centers.
- Radiation therapy near the device → device team plan, check after.
- Cardioversion: anterior–posterior pads, ≥8 cm from the generator; interrogate after.
- Central lines: guidewire in the RV can tangle new leads (<6 weeks) and trigger ICD sensing.
| Early (days–weeks) | Late |
|---|---|
| Pocket hematoma (esp. bridging heparin) | Device infection → full system extraction |
| Pneumothorax (subclavian access) | Lead fracture / insulation failure |
| Lead dislodgement → loss of capture | Tricuspid regurgitation from the RV lead |
| Perforation → tamponade (chest pain, hypotension) | Pacing-induced cardiomyopathy |
| Twiddler syndrome (patient rotates generator) | |
| Leadless: perforation and groin complications at implant; no pocket or leads to infect |
- Ask early: "If your heart stops at the end of a serious illness, do you want the device to shock you?"
- Deactivating shocks is legally and ethically withdrawing a treatment the patient no longer wants.
- Keep pacing on unless the patient asks (turning off pacing in a dependent patient needs a separate, careful discussion).
- At the bedside without a programmer: tape a magnet over the ICD to stop shocks in a dying patient.
Pitfalls
⌄
- Thinking a magnet turns off a pacemaker (it makes it pace all the time).
- Thinking a magnet changes pacing in an ICD (it does not — only tachy therapy).
- Putting a magnet on a leadless pacemaker and expecting a response.
- Missing hyperkalemia as the cause of loss of capture.
- Forgetting to re-enable ICD therapies after surgery.
- Sending a patient with multiple shocks home without interrogation.
Board corner
⌄
| Stem | Answer |
|---|---|
| Pacemaker, spikes without QRS, K 7.2 | Hyperkalemia → loss of capture → calcium |
| ICD fires repeatedly during AF with RVR, patient awake and talking | Inappropriate shocks → magnet + rate control |
| Hospice patient with ICD, wants no shocks | Deactivate tachy therapies (magnet if no programmer) |
| Regular paced tachycardia at upper rate after a PVC | Pacemaker-mediated tachycardia → magnet |
| New pacemaker, day 2, sudden hypotension and chest pain | Lead perforation → tamponade → echo |
| Elective laparoscopic cholecystectomy, ICD | Suspend tachy therapy (magnet or reprogram), pads on |
| EF 30%, LBBB 160 ms, NYHA III on GDMT | CRT-D |
References
⌄
- Kusumoto FM, et al. 2018 ACC/AHA/HRS bradycardia and conduction delay guideline. Circulation. 2019;140:e382.
- Al-Khatib SM, et al. 2017 AHA/ACC/HRS ventricular arrhythmias and SCD guideline. Circulation. 2018;138:e272.
- Crossley GH, et al. HRS/ASA expert consensus on perioperative management of CIEDs. Heart Rhythm. 2011;8:1114.
- ASA Practice Advisory for perioperative management of patients with CIEDs. Anesthesiology. 2020;132:225.
- Indik JH, et al. 2017 HRS expert consensus on MRI and radiation exposure in patients with CIEDs. Heart Rhythm. 2017;14:e97.
- Lampert R, et al. HRS consensus on CIED management at end of life. Heart Rhythm. 2010;7:1008.
- MGH White Book 2019–2020: Devices — PPM & ICD (structure).